13
6 RECM Alarm Test Element
6.1 Introduction
This test ensures that the ESU will sound an alarm if the resistance between the two electrodes exceeds a
specified limit. The QA-ES will gradually increase the resistance, starting at 10 Ohm and increasing through all
available load settings. At a certain value, the ESU should sound an alarm. It is this value that will be returned.
The user must define the acceptable alarm level through the expected results dialog. On this comparison the test
will either pass or fail.
6.2 Custom setup
6.2.1 Software Settings
The software settings are as section 5.2.
6.2.2 Hardware Settings
The hardware settings are found on the left of the custom setup dialog within the test settings box.
Delay
This is the time between the loads being applied to the ESU.
6.3 Expected Results
This performance test will contain the acceptable upper limit for the alarm to trigger.