1 Ansur QA-ES plug-in Version 1.4.4 User Manual

Strany 2 - Table of Contents

10 Use the Operand menu to select how Ansur shall calculate the limit given according to the power values defined in the test details. The operand c

Strany 3 - 1 Getting Started

11 5 H.F. Leakage Test Element 5.1 Introduction This test checks to see whether all electrodes on an ESU have leakage currents within acceptable lim

Strany 4 - 2.2 Test Element Definition

12 Active The number of active electrodes at the ESU. Neutral The number of neutral electrodes at the ESU. The second part is found on the right o

Strany 5 - 3.2 Custom setup

13 6 RECM Alarm Test Element 6.1 Introduction This test ensures that the ESU will sound an alarm if the resistance between the two electrodes exceed

Strany 6 - 3.3 Expected Results

14 7 Test Guide The test guide is the users interface with the QA-ES during the running of the test template. Click on User comment tab to enter a

Strany 7

15 Prior to each measurement within a test the software will display a prompt for the user to set the power, set the mode and set up the next electr

Strany 8 - 4 Output Test Element

16 Appendix A. QA-ES Performance Tests The following table lists all performance tests provided by the QA-ES plug-in. Performance test Unit Power

Strany 9 - 4.3 Expected Results

17 Appendix B. Retrieving updates Updates for Ansur are published from Fluke Biomedical’s web pages - http://www.flukebiomedical.com

Strany 10

18 Appendix C. References Ref. Title Author Version 1 QA-ES User & Service Manual Fluke Biomedical. 1.30-2 2 Ansur Test Executive User Man

Strany 11 - 5 H.F. Leakage Test Element

2 Table of Contents 1 Getting Started...

Strany 12 - 5.3 Expected Results

3 1 Getting Started 1.1 Introduction Ansur Test Automation Software is the evolution of the PRO-Soft software family and will become the backbone o

Strany 13 - 6 RECM Alarm Test Element

4 2 Introducing the QA-ES plug-in 2.1 Test elements After installing the QA-ES plug-in the QA-ES Performance tests will be added to Ansur’s Test Ex

Strany 14 - 7 Test Guide

5 3 Power Distribution Test Element 3.1 Introduction Within this test the user has the opportunity to determine the power performance of an ESU over

Strany 15 - 7.1 Toolbar buttons

6 The second part is found on the right of the custom setup dialog within the test details box. This determines the number and range of load mea

Strany 16 - QA-ES Performance Tests

7 So for the example below, each power performance measurement must be +/-20 W of its defined power value.

Strany 17 - Retrieving updates

8 4 Output Test Element 4.1 Introduction Within this test the user has the opportunity to determine the power performance of an ESU over a desired r

Strany 18 - References

9 Delay This is the delay between the changing of the loads and taking a power measurement. Mode Defines the mode that the ESU is to be in. The mod